CVEfocus › CVE database › CVE-2017-12714

CVE-2017-12714

EPSS 0.6% (higher than 51% of CVEs)Published 2018-04-25Updated 2024-09-17Assigner: icscert

Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted "RF wake-up" commands that can be received, which may allow a nearby attacker to repeatedly send commands to reduce pacemaker battery life. CVSS v3 base score: 5.3, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:N/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.

CWE-920

What to do

None of the Linux distributions or Windows releases we track list an affected package for this CVE. Check the vendor-reported products below.

Is your server affected? One read-only command checks every installed package against this CVE and all others, using each vendor's fixed versions. Get started or get a free API key.

curl -sS "https://api.cvefocus.com/v1/check?format=text" -H "Authorization: Bearer $CVEFOCUS_KEY" \
  -H "Content-Type: text/plain" --data-binary @inventory.txt

What the vendor says

As published in the CVE record by icscert, the CVE's assigner (usually the vendor). Check the vendor's advisory before acting.

References

About this page

Compiled by CVEfocus from the CVE List, the security data and VEX statements of Ubuntu, Debian, Red Hat, AlmaLinux, Rocky Linux and Microsoft, CISA's Known Exploited Vulnerabilities catalog and FIRST EPSS, refreshed several times a day (data status). Information only; see the disclaimer. Official record: cve.org · NVD.