CVE-2018-10631Medtronic N'Vision Clinician Programmer Protection Mechanism Failure
The 8840 Clinician Programmer executes the application program from the 8870 Application Card. An attacker with physical access to an 8870 Application Card and sufficient technical capability can modify the contents of this card, including the binary executables. If modified to bypass protection mechanisms, this malicious code will be run when the card is inserted into an 8840 Clinician Programmer.
What to do
None of the Linux distributions or Windows releases we track list an affected package for this CVE. Check the vendor-reported products below.
Is your server affected? One read-only command checks every installed package against this CVE and all others, using each vendor's fixed versions. Get started or get a free API key.
curl -sS "https://api.cvefocus.com/v1/check?format=text" -H "Authorization: Bearer $CVEFOCUS_KEY" \ -H "Content-Type: text/plain" --data-binary @inventory.txt
What the vendor says
As published in the CVE record by icscert, the CVE's assigner (usually the vendor). Check the vendor's advisory before acting.
- Medtronic 8840 N’Vision Clinician Programmer: affected: All versions
- Medtronic 8870 N’Vision removable Application Card: affected: All versions
Workaround: Medtronic has not developed a product update to address the vulnerabilities, but is reinforcing security reminders within this advisory to help reduce the risk associated with the vulnerabilities. The 8870 Therapy Application card stores PHI and PII as part of its normal operating procedure and should be handled, managed and secured in a manner consistent with the applicable laws for patient data privacy. Medtronic recommends users take additional defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, hospitals and clinicians should: * Maintain strict physical control of the 8870 application card. * Use only legitimately obtained 8870 cards and not cards provided by any third party as firmware and system updates are provided directly by Medtronic using new 8870 application cards. * 8840 Programmers and 8870 Therapy Application compact flash…
References
- https://global.medtronic.com/xg-en/product-security/security-bulletins/nvision.html
- https://www.medtronic.com/security
- https://ics-cert.us-cert.gov/advisories/ICSMA-18-137-01
- http://www.securityfocus.com/bid/104213
About this page
Compiled by CVEfocus from the CVE List, the security data and VEX statements of Ubuntu, Debian, Red Hat, AlmaLinux, Rocky Linux and Microsoft, CISA's Known Exploited Vulnerabilities catalog and FIRST EPSS, refreshed several times a day (data status). Information only; see the disclaimer. Official record: cve.org · NVD.