CVE-2018-18984Medtronic 9790, 2090 CareLink, and 29901 Encore Programmers Missing Encryption of…
Medtronic CareLink and Encore Programmers do not encrypt or do not sufficiently encrypt sensitive PII and PHI information while at rest .
What to do
None of the Linux distributions or Windows releases we track list an affected package for this CVE. Check the vendor-reported products below.
Is your server affected? One read-only command checks every installed package against this CVE and all others, using each vendor's fixed versions. Get started or get a free API key.
curl -sS "https://api.cvefocus.com/v1/check?format=text" -H "Authorization: Bearer $CVEFOCUS_KEY" \ -H "Content-Type: text/plain" --data-binary @inventory.txt
What the vendor says
As published in the CVE record by icscert, the CVE's assigner (usually the vendor). Check the vendor's advisory before acting.
- Medtronic CareLink 9790 Programmer: affected: All versions
- Medtronic CareLink 2090 Programmer: affected: All versions
- Medtronic 29901 Encore Programmer: affected: All versions
Workaround: The CareLink 9790 Programmer has been placed into end-of-life status and is no longer supported by Medtronic. Medtronic recommends users no longer use the 9790 for any purpose. The CareLink 2090 and 29901 Encore programmers store PHI and PII as part of their normal operating procedure. Medtronic recommends that when devices are storing PHI/PII it should be retained on these programmers for the least amount of time necessary, and should be handled, managed and secured in a manner consistent with the applicable laws for patient data privacy. Please contact a Medtronic representative for proper disposal and PHI/PII retention setting instructions. All affected programmers allow for the manual deletion of programmer-generated reports, which could contain PHI/PII. Medtronic recommends users delete these reports when no longer needed and prior to any disposition of the programmer. Medtronic…
References
- https://global.medtronic.com/xg-en/product-security/security-bulletins/carelink-9790-2090-29901.html
- https://ics-cert.us-cert.gov/advisories/ICSMA-18-347-01
- http://www.securityfocus.com/bid/106215
About this page
Compiled by CVEfocus from the CVE List, the security data and VEX statements of Ubuntu, Debian, Red Hat, AlmaLinux, Rocky Linux and Microsoft, CISA's Known Exploited Vulnerabilities catalog and FIRST EPSS, refreshed several times a day (data status). Information only; see the disclaimer. Official record: cve.org · NVD.