CVEfocus › CVE database › CVE-2018-8870

CVE-2018-8870Medtronic MyCareLink Patient Monitor Use of Hard-coded Password

CVSS 6.4 MediumEPSS 0.4% (higher than 31% of CVEs)Published 2018-07-02Updated 2025-05-22Assigner: icscert

Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains a hard-coded operating system password. An attacker with physical access can remove the case of the device, connect to the debug port, and use the password to gain privileged access to the operating system.

CWE-259

What to do

None of the Linux distributions or Windows releases we track list an affected package for this CVE. Check the vendor-reported products below.

Is your server affected? One read-only command checks every installed package against this CVE and all others, using each vendor's fixed versions. Get started or get a free API key.

curl -sS "https://api.cvefocus.com/v1/check?format=text" -H "Authorization: Bearer $CVEFOCUS_KEY" \
  -H "Content-Type: text/plain" --data-binary @inventory.txt

What the vendor says

As published in the CVE record by icscert, the CVE's assigner (usually the vendor). Check the vendor's advisory before acting.

Solution: Medtronic will release several rolling over-the-air product updates that will mitigate the vulnerabilities described within this advisory. These updates will be applied to devices automatically as part of standard, reoccurring update processes. In addition, Medtronic has increased security monitoring of affected devices and related infrastructure. Medtronic has released additional patient focused information, at the following location: https://www.medtronic.com/security

Workaround: Medtronic recommends users take additional defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should: * Maintain good physical controls over the home monitor as the best mitigation to these vulnerabilities. * Only use home monitors obtained directly from their healthcare provider or a Medtronic representative to ensure integrity of the system. * Report any concerning behavior regarding their home monitor to their healthcare provider or a Medtronic representative. Medtronic has released additional patient focused information, at the following location: https://www.medtronic.com/security

References

About this page

Compiled by CVEfocus from the CVE List, the security data and VEX statements of Ubuntu, Debian, Red Hat, AlmaLinux, Rocky Linux and Microsoft, CISA's Known Exploited Vulnerabilities catalog and FIRST EPSS, refreshed several times a day (data status). Information only; see the disclaimer. Official record: cve.org · NVD.