CVE-2019-10964Medtronic MiniMed 508 and Paradigm Series Insulin Pumps Improper Access Control

CVSS 7.1 HighEPSS 1.2% (higher than 70% of CVEs)Published 2019-06-28Updated 2025-05-22Assigner: icscert

Medtronic MiniMed Insulin Pumps are designed to communicate using a wireless RF with other devices, such as blood glucose meters, glucose sensor transmitters, and CareLink USB devices. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with adjacent access to one of the affected insulin pump models can inject, replay, modify, and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.

CWE-284

What to do

None of the Linux distributions or Windows releases we track list an affected package for this CVE. Check the vendor-reported products below.

Is your server affected? One read-only command checks every installed package against this CVE and all others, using each vendor's fixed versions. Get started or get a free API key.

curl -sS "https://api.cvefocus.com/v1/check?format=text" -H "Authorization: Bearer $CVEFOCUS_KEY" \
  -H "Content-Type: text/plain" --data-binary @inventory.txt

What the vendor says

As published in the CVE record by icscert, the CVE's assigner (usually the vendor). Check the vendor's advisory before acting.

Workaround: Medtronic recommends U.S. patients who are currently using the affected products talk to their healthcare provider about changing to a newer model insulin pump with increased cybersecurity protection. Patients outside the U.S. will receive a notification letter with instructions based on the country where they live. Medtronic recommends all patients take the cybersecurity precautions indicated below. CYBERSECURITY PRECAUTIONS RECOMMENDED FOR ALL PATIENTS: * Maintain tight physical control of the pump and devices connected to the pump * Do not share pump serial number * Be attentive to pump notifications, alarms, and alerts * Immediately cancel any unintended boluses (a single dose of insulin administered all at once) * Do not connect to any third-party devices or use any software not authorized by Medtronic * Disconnect CareLink USB devices from computers when not being used to…

References

About this page

Compiled by CVEfocus from the CVE List, the security data and VEX statements of Ubuntu, Debian, Red Hat, AlmaLinux, Rocky Linux and Microsoft, CISA's Known Exploited Vulnerabilities catalog and FIRST EPSS, refreshed several times a day (data status). Information only; see the disclaimer. Official record: cve.org · NVD.