CVE-2019-10964Medtronic MiniMed 508 and Paradigm Series Insulin Pumps Improper Access Control
Medtronic MiniMed Insulin Pumps are designed to communicate using a wireless RF with other devices, such as blood glucose meters, glucose sensor transmitters, and CareLink USB devices. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with adjacent access to one of the affected insulin pump models can inject, replay, modify, and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.
What to do
None of the Linux distributions or Windows releases we track list an affected package for this CVE. Check the vendor-reported products below.
Is your server affected? One read-only command checks every installed package against this CVE and all others, using each vendor's fixed versions. Get started or get a free API key.
curl -sS "https://api.cvefocus.com/v1/check?format=text" -H "Authorization: Bearer $CVEFOCUS_KEY" \ -H "Content-Type: text/plain" --data-binary @inventory.txt
What the vendor says
As published in the CVE record by icscert, the CVE's assigner (usually the vendor). Check the vendor's advisory before acting.
- Medtronic MiniMed 508 pump: affected: All versions
- Medtronic MiniMed Paradigm 511 pump: affected: All versions
- Medtronic MiniMed Paradigm 512/712 pumps: affected: All versions
- Medtronic MiniMed Paradigm 712E pump: affected: All versions
- Medtronic MiniMed Paradigm 515/715 pumps: affected: All versions
- Medtronic MiniMed Paradigm 522/722 pumps: affected: All versions
- Medtronic MiniMed Paradigm 522K/722K pumps: affected: All versions
- Medtronic MiniMed Paradigm 523/723 pumps: affected: up to Software Versions 2.4A
- Medtronic MiniMed Paradigm 523K/723K pumps: affected: up to Software Versions 2.4A
- Medtronic MiniMed Paradigm Veo 554/754 pumps: affected: up to Software Versions 2.6A
- Medtronic MiniMed Paradigm Veo 554CM/754CM pumps: affected: up to Software Versions 2.7A
Workaround: Medtronic recommends U.S. patients who are currently using the affected products talk to their healthcare provider about changing to a newer model insulin pump with increased cybersecurity protection. Patients outside the U.S. will receive a notification letter with instructions based on the country where they live. Medtronic recommends all patients take the cybersecurity precautions indicated below. CYBERSECURITY PRECAUTIONS RECOMMENDED FOR ALL PATIENTS: * Maintain tight physical control of the pump and devices connected to the pump * Do not share pump serial number * Be attentive to pump notifications, alarms, and alerts * Immediately cancel any unintended boluses (a single dose of insulin administered all at once) * Do not connect to any third-party devices or use any software not authorized by Medtronic * Disconnect CareLink USB devices from computers when not being used to…
References
- https://global.medtronic.com/xg-en/product-security/security-bulletins/minimed-508-paradigm.html
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-19-178-01
- http://www.securityfocus.com/bid/108926
About this page
Compiled by CVEfocus from the CVE List, the security data and VEX statements of Ubuntu, Debian, Red Hat, AlmaLinux, Rocky Linux and Microsoft, CISA's Known Exploited Vulnerabilities catalog and FIRST EPSS, refreshed several times a day (data status). Information only; see the disclaimer. Official record: cve.org · NVD.