CVE-2019-13539Medtronic Valleylab FT10 and FX8 Reversible One-way Hash
Medtronic Valleylab Exchange Client version 3.4 and below, Valleylab FT10 Energy Platform (VLFT10GEN) software version 4.0.0 and below, and Valleylab FX8 Energy Platform (VLFX8GEN) software version 1.1.0 and below use the descrypt algorithm for OS password hashing. While interactive, network-based logons are disabled, and attackers can use the other vulnerabilities within this report to obtain local shell access and access these hashes.
What to do
None of the Linux distributions or Windows releases we track list an affected package for this CVE. Check the vendor-reported products below.
Is your server affected? One read-only command checks every installed package against this CVE and all others, using each vendor's fixed versions. Get started or get a free API key.
curl -sS "https://api.cvefocus.com/v1/check?format=text" -H "Authorization: Bearer $CVEFOCUS_KEY" \ -H "Content-Type: text/plain" --data-binary @inventory.txt
What the vendor says
As published in the CVE record by icscert, the CVE's assigner (usually the vendor). Check the vendor's advisory before acting.
- Medtronic Valleylab Exchange Client: affected: up to 3.4
- Medtronic Valleylab FT10 Energy Platform (VLFT10GEN): affected: up to software version 4.0.0
- Medtronic Valleylab FX8 Energy Platform (VLFX8GEN): affected: up to software version 1.1.0
Solution: Software patches are currently available for the FT10 platform and will be available in early 2020 for the FX8 platform. Until these updates can be applied, Medtronic recommends to either disconnect affected products from IP networks or to segregate those networks, such that the devices are not accessible from an untrusted network (e.g., Internet). Patches can be downloaded at the following location: https://www.medtronic.com/covidien/en-us/support/software.html Medtronic has released additional patient focused information, at the following location: https://www.medtronic.com/security
References
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-19-311-02
- https://global.medtronic.com/xg-en/product-security/security-bulletins/valleylab-generator-rfid-vulnerabilities.html
About this page
Compiled by CVEfocus from the CVE List, the security data and VEX statements of Ubuntu, Debian, Red Hat, AlmaLinux, Rocky Linux and Microsoft, CISA's Known Exploited Vulnerabilities catalog and FIRST EPSS, refreshed several times a day (data status). Information only; see the disclaimer. Official record: cve.org · NVD.