CVEfocus › CVE database › CVE-2020-12023

CVE-2020-12023Philips IntelliBridge Enterprise IBE Insertion of Sensitive Information into Log File

CVSS 2.0 LowEPSS 0.5% (higher than 37% of CVEs)Published 2020-06-11Updated 2025-06-04Assigner: icscert

Philips IntelliBridge Enterprise (IBE), Versions B.12 and prior, IntelliBridge Enterprise system integration with SureSigns (VS4), EarlyVue (VS30) and IntelliVue Guardian (IGS). Unencrypted user credentials received in the IntelliBridge Enterprise (IBE) are logged within the transaction logs, which are secured behind the login based administrative web portal. The unencrypted user credentials sent from the affected products listed above, for the purpose of handshake or authentication with the Enterprise Systems, are logged as the payload in IntelliBridge Enterprise (IBE) within the transaction logs. An attacker with administrative privileges could exploit this vulnerability to read plain text credentials from log files.

CWE-532

What to do

None of the Linux distributions or Windows releases we track list an affected package for this CVE. Check the vendor-reported products below.

Is your server affected? One read-only command checks every installed package against this CVE and all others, using each vendor's fixed versions. Get started or get a free API key.

curl -sS "https://api.cvefocus.com/v1/check?format=text" -H "Authorization: Bearer $CVEFOCUS_KEY" \
  -H "Content-Type: text/plain" --data-binary @inventory.txt

What the vendor says

As published in the CVE record by icscert, the CVE's assigner (usually the vendor). Check the vendor's advisory before acting.

Solution: Philips plans a new release (IBE B.13) by the end of Q4 2020 that remediates the security risk by not logging the plain text user credentials in the log file. Users with questions regarding their specific Philips IntelliBridge Enterprise installation should contact their local Philips service support team, or regional service support. Users can see the Philips advisory http://www.philips.com/productsecurity for more details, and contact the Philips Customer Service Solutions https://www.usa.philips.com/healthcare/solutions/customer-service-solutions team for additional guidance. Please see the Philips product security website https://www.philips.com/productsecurity for the latest security information for Philips products.

Workaround: As an interim mitigation to this vulnerability, Philips recommends the following: * The IBE transaction logs are only accessible with administrative privileges. An additional account can be created on the IBE system with limited privileges, for service engineers. * Reduce log retention to an acceptable timeframe that allows for recovery activities. Users with questions regarding their specific Philips IntelliBridge Enterprise installation should contact their local Philips service support team, or regional service support. Users can see the Philips advisory http://www.philips.com/productsecurity for more details, and contact the Philips Customer Service Solutions https://www.usa.philips.com/healthcare/solutions/customer-service-solutions team for additional guidance. Please see the Philips product security website https://www.philips.com/productsecurity for the latest security…

References

About this page

Compiled by CVEfocus from the CVE List, the security data and VEX statements of Ubuntu, Debian, Red Hat, AlmaLinux, Rocky Linux and Microsoft, CISA's Known Exploited Vulnerabilities catalog and FIRST EPSS, refreshed several times a day (data status). Information only; see the disclaimer. Official record: cve.org · NVD.