CVEfocus › CVE database › CVE-2020-16212

CVE-2020-16212Philips Patient Monitoring Devices Exposure of Resource to Wrong Sphere

EPSS 0.4% (higher than 23% of CVEs)Published 2020-09-11Updated 2024-08-04Assigner: icscert

In Patient Information Center iX (PICiX) Versions B.02, C.02, C.03, the product exposes a resource to the wrong control sphere, providing unintended actors with inappropriate access to the resource. The application on the surveillance station operates in kiosk mode, which is vulnerable to local breakouts that could allow an attacker with physical access to escape the restricted environment with limited privileges.

CWE-668

What to do

None of the Linux distributions or Windows releases we track list an affected package for this CVE. Check the vendor-reported products below.

Is your server affected? One read-only command checks every installed package against this CVE and all others, using each vendor's fixed versions. Get started or get a free API key.

curl -sS "https://api.cvefocus.com/v1/check?format=text" -H "Authorization: Bearer $CVEFOCUS_KEY" \
  -H "Content-Type: text/plain" --data-binary @inventory.txt

What the vendor says

As published in the CVE record by icscert, the CVE's assigner (usually the vendor). Check the vendor's advisory before acting.

Solution: Philips released the following versions to remediate reported vulnerabilities: * Patient Information Center iX (PICiX) Version C.03 * Certificate revocation within the system was implemented for PIC iX and Performance Bridge FocalPoint in 2023. The implementation of the IntelliVue Patient Monitors will be completed in Q3 of 2024.

Workaround: As a mitigation to these vulnerabilities, Philips recommends the following: * The Philips patient monitoring network is required to be physically or logically isolated from the hospital local area network (LAN). Philips recommends using a firewall or routers that can implement access control lists restricting access in and out of the patient monitoring network for only necessary ports and IP addresses. Refer to the Philips Patient Monitoring System Security for Clinical Networks guide for additional information on [InCenter]( https://incenter.medical.philips.com/) https://incenter.medical.philips.com/) . * By default, the simple certificate enrollment protocol (SCEP) service is not running. When needed, the service is configured to run based on the duration or the number of certificates to be assigned. One certificate is default, but if a certificate is not issued, the service will…

References

About this page

Compiled by CVEfocus from the CVE List, the security data and VEX statements of Ubuntu, Debian, Red Hat, AlmaLinux, Rocky Linux and Microsoft, CISA's Known Exploited Vulnerabilities catalog and FIRST EPSS, refreshed several times a day (data status). Information only; see the disclaimer. Official record: cve.org · NVD.