CVE-2020-16228Philips Patient Monitoring Devices Improper Check for Certificate Revocation
In Patient Information Center iX (PICiX) Versions C.02 and C.03, PerformanceBridge Focal Point Version A.01, IntelliVue patient monitors MX100, MX400-MX550, MX750, MX850, and IntelliVue X3 Versions N and prior, the software does not check or incorrectly checks the revocation status of a certificate, which may cause it to use a compromised certificate.
What to do
None of the Linux distributions or Windows releases we track list an affected package for this CVE. Check the vendor-reported products below.
Is your server affected? One read-only command checks every installed package against this CVE and all others, using each vendor's fixed versions. Get started or get a free API key.
curl -sS "https://api.cvefocus.com/v1/check?format=text" -H "Authorization: Bearer $CVEFOCUS_KEY" \ -H "Content-Type: text/plain" --data-binary @inventory.txt
What the vendor says
As published in the CVE record by icscert, the CVE's assigner (usually the vendor). Check the vendor's advisory before acting.
- Philips Patient Information Center iX (PICiX): affected: C.02, C.03
- Philips PerformanceBridge Focal Point: affected: A.01
- Philips IntelliVue patient monitors: affected: MX100, MX400-MX550, MX750, MX850
- Philips IntelliVue X3: affected: up to N
Solution: Philips released the following versions to remediate reported vulnerabilities: * Patient Information Center iX (PICiX) Version C.03 * PerformanceBridge Focal Point * IntelliVue Patient Monitors Versions N.00 and N.01 * IntelliVue Patient Monitors Version M.04: Contact a Philips service support team https://www.usa.philips.com/healthcare/solutions/customer-service-solutions for an upgrade path * Certificate revocation within the system was implemented for PIC iX and Performance Bridge FocalPoint in 2023. The implementation of the IntelliVue Patient Monitors will be completed in Q3 of 2024.
Workaround: As a mitigation to these vulnerabilities, Philips recommends the following: * The Philips patient monitoring network is required to be physically or logically isolated from the hospital local area network (LAN). Philips recommends using a firewall or routers that can implement access control lists restricting access in and out of the patient monitoring network for only necessary ports and IP addresses. Refer to the Philips Patient Monitoring System Security for Clinical Networks guide for additional information on InCenter https://incenter.medical.philips.com/ . * By default, the simple certificate enrollment protocol (SCEP) service is not running. When needed, the service is configured to run based on the duration or the number of certificates to be assigned. One certificate is default, but if a certificate is not issued, the service will continue to run. Limit exposure by ensuring…
References
About this page
Compiled by CVEfocus from the CVE List, the security data and VEX statements of Ubuntu, Debian, Red Hat, AlmaLinux, Rocky Linux and Microsoft, CISA's Known Exploited Vulnerabilities catalog and FIRST EPSS, refreshed several times a day (data status). Information only; see the disclaimer. Official record: cve.org · NVD.