CVEfocus › CVE database › CVE-2020-25183

CVE-2020-25183Medtronic MyCareLink Smart Improper Authentication

CVSS 8.0 HighEPSS 0.8% (higher than 53% of CVEs)Published 2020-12-14Updated 2025-05-22Assigner: icscert

Medtronic MyCareLink Smart 25000 contains an authentication protocol vulnerability where the method used to authenticate between the MCL Smart Patient Reader and the Medtronic MyCareLink Smart mobile app is vulnerable to bypass. This vulnerability enables an attacker to use another mobile device or malicious application on the patient’s smartphone to authenticate to the patient’s Medtronic Smart Reader, fooling the device into believing it is communicating with the original Medtronic smart phone application when executed within range of Bluetooth communication.

CWE-287

What to do

None of the Linux distributions or Windows releases we track list an affected package for this CVE. Check the vendor-reported products below.

Is your server affected? One read-only command checks every installed package against this CVE and all others, using each vendor's fixed versions. Get started or get a free API key.

curl -sS "https://api.cvefocus.com/v1/check?format=text" -H "Authorization: Bearer $CVEFOCUS_KEY" \
  -H "Content-Type: text/plain" --data-binary @inventory.txt

What the vendor says

As published in the CVE record by icscert, the CVE's assigner (usually the vendor). Check the vendor's advisory before acting.

Solution: A firmware update to eliminates these vulnerabilities has been developed by Medtronic and is available by updating the MyCareLink Smartapp via the associated mobile application store. Upgrading to the latest v5.2 mobile application version will ensure the Patient Reader is also updated on next use. The user’s smart phone must be updated to the following operating system version for the patches to be applied: iOS 10 and above; Android 6.0 and above. Medtronic has released additional patient focused information https://www.medtronic.com/security : https://www.medtronic.com/xg-en/product-security/security-bulletins.html https://www.medtronic.com/xg-en/product-security/security-bulletins.html

Workaround: In response to these vulnerabilities, Medtronic has applied additional controls for monitoring and responding to improper use of the MCL Smart Patient Reader: * Medtronic has implemented enhanced integrity validation (EIV) technology, which provides early detection and real-time mitigation of known vulnerability exploitation attempts. * Medtronic has also implemented advanced detection system technology, which enables device-level logging and monitoring of all device activity and behavior. Medtronic recommends that users take additional defensive measures to minimize risk. Specifically, users should: * Maintain good physical control over home monitors. * This includes only using home monitors in private environments such as a home, apartment, or otherwise physically controlled environment. * Use only home monitors obtained directly from your healthcare provider or a Medtronic…

References

About this page

Compiled by CVEfocus from the CVE List, the security data and VEX statements of Ubuntu, Debian, Red Hat, AlmaLinux, Rocky Linux and Microsoft, CISA's Known Exploited Vulnerabilities catalog and FIRST EPSS, refreshed several times a day (data status). Information only; see the disclaimer. Official record: cve.org · NVD.