CVEfocus › CVE database › CVE-2021-23233

CVE-2021-23233Fresenius Kabi Agilia Connect Infusion System

CVSS 7.3 HighEPSS 1.0% (higher than 62% of CVEs)Published 2022-01-21Updated 2025-04-16Assigner: icscert

Sensitive endpoints in Fresenius Kabi Agilia Link+ v3.0 and prior can be accessed without any authentication information such as the session cookie. An attacker can send requests to sensitive endpoints as an unauthenticated user to perform critical actions or modify critical configuration parameters.

CWE-284

What to do

None of the Linux distributions or Windows releases we track list an affected package for this CVE. Check the vendor-reported products below.

Is your server affected? One read-only command checks every installed package against this CVE and all others, using each vendor's fixed versions. Get started or get a free API key.

curl -sS "https://api.cvefocus.com/v1/check?format=text" -H "Authorization: Bearer $CVEFOCUS_KEY" \
  -H "Content-Type: text/plain" --data-binary @inventory.txt

What the vendor says

As published in the CVE record by icscert, the CVE's assigner (usually the vendor). Check the vendor's advisory before acting.

Solution: Fresenius Kabi has created new versions to address these vulnerabilities: • Link+ v3.0 (D16 or later) • VSS v1.0.3 (or later) • Agilia Connect Pumps Wifi Module (D29 or later) • Agilia Connect Partner v3.3.2 (or later) Fresenius Kabi has initiated communication on this topic in April 2021 with customers to inform them about availability of the new versions in corresponding countries. Contact Fresenius Kabi online or by phone at 1-800-333-6925 for more information on updating to the newest version. Fresenius Kabi also identified that early Link+ devices (approximatively 1200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in customers’ installation, Fresenius Kabi recommends users to rely on CISA recommendation.

Workaround: CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should: • Minimize network exposure for all control system devices and/or systems, and ensure they are not accessible from the Internet. • Locate control system networks and remote devices behind firewalls and isolate them from the business network. • When remote access is required, use secure methods, such as virtual private networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as its connected devices.

References

About this page

Compiled by CVEfocus from the CVE List, the security data and VEX statements of Ubuntu, Debian, Red Hat, AlmaLinux, Rocky Linux and Microsoft, CISA's Known Exploited Vulnerabilities catalog and FIRST EPSS, refreshed several times a day (data status). Information only; see the disclaimer. Official record: cve.org · NVD.