CVEfocus › CVE database › CVE-2021-38392

CVE-2021-38392Improper Access Control for Boston Scientific Zoom Latitude

CVSS 6.5 MediumEPSS 0.3% (higher than 11% of CVEs)Published 2021-10-04Updated 2024-09-17Assigner: icscert

A skilled attacker with physical access to the affected device can gain access to the hard disk drive of the device to change the telemetry region and could use this setting to interrogate or program an implantable device in any region in the world.

CWE-284

What to do

None of the Linux distributions or Windows releases we track list an affected package for this CVE. Check the vendor-reported products below.

Is your server affected? One read-only command checks every installed package against this CVE and all others, using each vendor's fixed versions. Get started or get a free API key.

curl -sS "https://api.cvefocus.com/v1/check?format=text" -H "Authorization: Bearer $CVEFOCUS_KEY" \
  -H "Content-Type: text/plain" --data-binary @inventory.txt

What the vendor says

As published in the CVE record by icscert, the CVE's assigner (usually the vendor). Check the vendor's advisory before acting.

Workaround: Boston Scientific is in the process of transitioning all users to a replacement programmer with enhanced security, the LATITUDE Programming System, Model 3300. Boston Scientific will not issue a product update to address the identified vulnerabilities in the ZOOM LATITUDE Programming System, Model 3120.

References

About this page

Compiled by CVEfocus from the CVE List, the security data and VEX statements of Ubuntu, Debian, Red Hat, AlmaLinux, Rocky Linux and Microsoft, CISA's Known Exploited Vulnerabilities catalog and FIRST EPSS, refreshed several times a day (data status). Information only; see the disclaimer. Official record: cve.org · NVD.