CVE-2021-4479Dräger Atlan A350 1.00 <= 1.01 DoS via Medibus Interface
Dräger Atlan A350 versions 1.00 up to and including 1.01 contains an improper input handling vulnerability that allows attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can transmit malformed data to overload the internal processor, gradually disrupting device operation over several hours and causing loss of data transmission, delayed display of real-time curves, and deviation between displayed airway pressure values and screen curves.
What to do
None of the Linux distributions or Windows releases we track list an affected package for this CVE. Check the vendor-reported products below.
Is your server affected? One read-only command checks every installed package against this CVE and all others, using each vendor's fixed versions. Get started or get a free API key.
curl -sS "https://api.cvefocus.com/v1/check?format=text" -H "Authorization: Bearer $CVEFOCUS_KEY" \ -H "Content-Type: text/plain" --data-binary @inventory.txt
What the vendor says
As published in the CVE record by VulnCheck, the CVE's assigner (usually the vendor). Check the vendor's advisory before acting.
- Dräger Atlan A350: affected: 1.00 to 1.01
Vendor advisories:
- https://static.draeger.com/security/download/2021-05-04-PSA-21-124-1-Security-Advisory-At… (vendor-advisory)
References
- https://static.draeger.com/security/download/2021-05-04-PSA-21-124-1-Security-Advisory-Atlan.pdf
- https://www.vulncheck.com/advisories/drager-atlan-a350-dos-via-medibus-interface
About this page
Compiled by CVEfocus from the CVE List, the security data and VEX statements of Ubuntu, Debian, Red Hat, AlmaLinux, Rocky Linux and Microsoft, CISA's Known Exploited Vulnerabilities catalog and FIRST EPSS, refreshed several times a day (data status). Information only; see the disclaimer. Official record: cve.org · NVD.