CVE-2022-26388Use of Hard-Coded Password Vulnerability in ELI Electrocardiograph Devices
A use of hard-coded password vulnerability may allow authentication abuse.This issue affects ELI 380 Resting Electrocardiograph: Versions 2.6.0 and prior; ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph: Versions 2.3.1 and prior; ELI 250c/BUR 250c Resting Electrocardiograph: Versions 2.1.2 and prior; ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph: Versions 2.2.0 and prior.
What to do
Fixed builds are available from Alpine: see the table below for the fixed version per release.
Is your server affected? One read-only command checks every installed package against this CVE and all others, using each vendor's fixed versions. Get started or get a free API key.
curl -sS "https://api.cvefocus.com/v1/check?format=text" -H "Authorization: Bearer $CVEFOCUS_KEY" \ -H "Content-Type: text/plain" --data-binary @inventory.txt
Vendor status at a glance
Status by vendor and release
From each vendor's own security data and VEX statements. "Fixed" means a fixed build is published for that release.
| Release | Package | Status | Fixed version | Advisory |
|---|---|---|---|---|
| Alpine Linux 3.23 | thunderbird | Fixed | 91.7.0-r0 | ALPINE-CVE-2022-26388 |
| Alpine Linux 3.22 | thunderbird | Fixed | 91.7.0-r0 | ALPINE-CVE-2022-26388 |
| Alpine Linux 3.21 | thunderbird | Fixed | 91.7.0-r0 | ALPINE-CVE-2022-26388 |
| Alpine Linux 3.20 | thunderbird | Fixed | 91.7.0-r0 | ALPINE-CVE-2022-26388 |
| Alpine Linux 3.19 | thunderbird | Fixed | 91.7.0-r0 | ALPINE-CVE-2022-26388 |
| Alpine Linux 3.18 | thunderbird | Fixed | 91.7.0-r0 | ALPINE-CVE-2022-26388 |
| Alpine Linux 3.17 | thunderbird | Fixed | 91.7.0-r0 | ALPINE-CVE-2022-26388 |
| Alpine Linux 3.16 | thunderbird | Fixed | 91.7.0-r0 | ALPINE-CVE-2022-26388 |
What the vendor says
As published in the CVE record by Baxter, the CVE's assigner (usually the vendor). Check the vendor's advisory before acting.
- Welch Allyn ELI 380 Resting Electrocardiograph: affected: up to 2.6.0
- Welch Allyn ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph: affected: up to 2.3.1
- Welch Allyn ELI 250c/BUR 250c Resting Electrocardiograph: affected: up to 2.1.2
- Welch Allyn ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph: affected: up to 2.2.0
Solution: Hillrom has released software updates for all impacted devices to address these vulnerabilities. New product versions that mitigate these vulnerabilities are available as follows: * Welch Allyn ELI 380 Resting Electrocardiograph: available Q4 2023 * Welch Allyn ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph: available May 2022 * Welch Allyn ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph: available Q4 2023 Hillrom recommends users upgrade to the latest product versions. Information on how to update these products can be found on the Hillrom disclosure page https://hillrom.com/en/responsible-disclosures/ .
Workaround: Hillrom recommends the following workarounds to help reduce risk: * Apply proper network and physical security controls. * Ensure a unique encryption key is configured for ELI Link and Cardiograph. * Where possible, use a firewall to prevent communication on Port 21 FTP service, Port 22 SSH (Secure Shell Connection), and Port 23 Telnet service.
References
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-22-167-01
- https://hillrom.com/en/responsible-disclosures/
About this page
Compiled by CVEfocus from the CVE List, the security data and VEX statements of Ubuntu, Debian, Red Hat, AlmaLinux, Rocky Linux and Microsoft, CISA's Known Exploited Vulnerabilities catalog and FIRST EPSS, refreshed several times a day (data status). Information only; see the disclaimer. Official record: cve.org · NVD.