CVEfocus › CVE database › CVE-2022-36385

CVE-2022-36385Contec Health CMS8000

CVSS 6.8 MediumEPSS 0.4% (higher than 30% of CVEs)Published 2022-09-13Updated 2025-04-16Assigner: icscert

A threat actor with momentary access to the device can plug in a USB drive and perform a malicious firmware update, resulting in permanent changes to device functionality. No authentication or controls are in place to prevent a threat actor from maliciously modifying firmware and performing a drive-by attack to load the firmware on any CMS8000 device.

CWE-284

What to do

None of the Linux distributions or Windows releases we track list an affected package for this CVE. Check the vendor-reported products below.

Is your server affected? One read-only command checks every installed package against this CVE and all others, using each vendor's fixed versions. Get started or get a free API key.

curl -sS "https://api.cvefocus.com/v1/check?format=text" -H "Authorization: Bearer $CVEFOCUS_KEY" \
  -H "Content-Type: text/plain" --data-binary @inventory.txt

What the vendor says

As published in the CVE record by icscert, the CVE's assigner (usually the vendor). Check the vendor's advisory before acting.

Workaround: Contec Health has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Contec Health for additional information. The following mitigations could assist in reducing the risk for exploitation of vulnerabilities: Disabling UART functionality at the CPU level Enforcing unique device authentication before granting access to the terminal / bootloader Where possible, enforcing secure boot. Tamper stickers on the device casing to indicate when a device has been opened

References

About this page

Compiled by CVEfocus from the CVE List, the security data and VEX statements of Ubuntu, Debian, Red Hat, AlmaLinux, Rocky Linux and Microsoft, CISA's Known Exploited Vulnerabilities catalog and FIRST EPSS, refreshed several times a day (data status). Information only; see the disclaimer. Official record: cve.org · NVD.