CVE-2022-43557BD BodyGuard™ Pumps – RS-232 Interface Vulnerability
The BD BodyGuard™ infusion pumps specified allow for access through the RS-232 (serial) port interface. If exploited, threat actors with physical access, specialized equipment and knowledge may be able to configure or disable the pump. No electronic protected health information (ePHI), protected health information (PHI) or personally identifiable information (PII) is stored in the pump.
What to do
None of the Linux distributions or Windows releases we track list an affected package for this CVE. Check the vendor-reported products below.
Is your server affected? One read-only command checks every installed package against this CVE and all others, using each vendor's fixed versions. Get started or get a free API key.
curl -sS "https://api.cvefocus.com/v1/check?format=text" -H "Authorization: Bearer $CVEFOCUS_KEY" \ -H "Content-Type: text/plain" --data-binary @inventory.txt
What the vendor says
As published in the CVE record by BD, the CVE's assigner (usually the vendor). Check the vendor's advisory before acting.
- Becton, Dickinson and Company (BD) BodyGuard™ Pump: affected: BD BodyGuard™ , CME BodyGuard™ 323 (2nd Edition), CME BodyGuard™ 323 Color Vision (2nd Edition), CME BodyGuard™ 323 Color Vision (3rd Edition), CME BodyGuard™ Twins (2nd Edition)
Workaround: BD recommends the following mitigations and compensating controls to reduce risk associated with this vulnerability: • Ensure physical access controls are in place and only authorized end-users have access to BD BodyGuard™ pumps. • Ensure that only BD-approved equipment is connected to the RS-232 interface of the affected pumps. • Ensure that no equipment is connected to the RS-232 interface when the affected pumps are delivering infusions. • Protect connected computer systems with BodyComm™ software with standard security measures.
Vendor advisories:
- https://www.bd.com/en-us/about-bd/cybersecurity/bulletin/bd-bodyguard-pumps-rs-232-interf… (vendor-advisory)
References
About this page
Compiled by CVEfocus from the CVE List, the security data and VEX statements of Ubuntu, Debian, Red Hat, AlmaLinux, Rocky Linux and Microsoft, CISA's Known Exploited Vulnerabilities catalog and FIRST EPSS, refreshed several times a day (data status). Information only; see the disclaimer. Official record: cve.org · NVD.