CVEfocus › CVE database › CVE-2026-17583

CVE-2026-17583Thermo Fisher Applied Biosystems Genetic Analyzers Missing Support for Integrity Check

CVSS 8.4 HighEPSS 0.1% (higher than 2% of CVEs)Published 2026-08-05Updated 2026-08-06Assigner: icscert

The affected Thermo Fisher Applied Biosystems Genetic Analyzers are vulnerable because .fsa/.hid output files can be edited. An attacker could tamper with these files, altering DNA data and resulting in inaccurate DNA test outcomes.

CWE-353

What to do

Ubuntu has not shipped a fix for every release yet.

Is your server affected? One read-only command checks every installed package against this CVE and all others, using each vendor's fixed versions. Get started or get a free API key.

curl -sS "https://api.cvefocus.com/v1/check?format=text" -H "Authorization: Bearer $CVEFOCUS_KEY" \
  -H "Content-Type: text/plain" --data-binary @inventory.txt

Vendor status at a glance

3Fix pending
3Under investigation

Status by vendor and release

From each vendor's own security data and VEX statements. "Fixed" means a fixed build is published for that release.

ReleasePackageStatusFixed versionAdvisory
Ubuntu 26.10geneticFix pendingUBUNTU-CVE-2026-17583
Ubuntu 26.04geneticUnder investigationUBUNTU-CVE-2026-17583
Ubuntu 24.04geneticUnder investigationUBUNTU-CVE-2026-17583
Ubuntu 22.04geneticUnder investigationUBUNTU-CVE-2026-17583
Ubuntu 18.04geneticFix pendingUBUNTU-CVE-2026-17583
Ubuntu 16.04geneticFix pendingUBUNTU-CVE-2026-17583

What the vendor says

As published in the CVE record by icscert, the CVE's assigner (usually the vendor). Check the vendor's advisory before acting.

Solution: Thermo Fisher has developed security updates to address the vulnerability. The security updates implement the use of digital signatures on the instrument software that adds an extralayer of protection. Moving forward, this will help users verify that data files have not been modified. Applied Biosystems 3500/3500xL Series Data Collection Software: Update to version 4.0.3 https://downloads.thermofisher.com/3500_DCS_v4.0.3_Patch/v4.0.3_Patch_Installer.exe Applied Biosystems 3730/3730xL Series Data Collection Software: Update to version 5.0.3 https://downloads.thermofisher.com/3730xl_UDC_v5.0.3_Patch/3730xl_UDC_v5.0.3_Patch.exe Applied Biosystems SeqStudio Genetic Analyzer Data Collection Software: Update to version 1.2.6 https://downloads.thermofisher.com/SeqStudio/1.2.6/SeqStudio-1.2.6.abpkg Applied Biosystems SeqStudio Flex Series Instrument Software: Update to version 1.2.1…

Workaround: Applied Biosystems 3130 Series Data Collection Software: Product is End of Life (EoL), no update provided ABI PRISM 3100/3100-Avant Data Collection Software: Product is End of Life (EoL), no update provided ABI PRISM 310 Data Collection Software: Product is End of Life (EoL), no update provided For users who are unable to immediately implement all applicable security updates, Thermo Fisher Scientific recommends implementing the following interim mitigation measures until the applicable updates have been installed: * Maintain a secure chain of custody for files generated by the HID instrumentation throughout the analysis workflow. * Store generated files on encrypted, password-protected storage media (for example, encrypted USB drives or encrypted hard drives). * Restrict access to generated files to authorized personnel in accordance with your laboratory's access control policies. *…

References

About this page

Compiled by CVEfocus from the CVE List, the security data and VEX statements of Ubuntu, Debian, Red Hat, AlmaLinux, Rocky Linux and Microsoft, CISA's Known Exploited Vulnerabilities catalog and FIRST EPSS, refreshed several times a day (data status). Information only; see the disclaimer. Official record: cve.org · NVD.